| Package | Affected versions | Patched versions |
|---|---|---|
| abi prism 3100/3100-avant data collection software | ≤ 2.0 | |
| abi prism 310 data collection software | ≤ 3.1 | |
| applied biosystems 3130 series data collection software | ≤ 4.1 | |
| applied biosystems 3500/3500xl series data collection software | < 4.0.3 | 4.0.3 |
| applied biosystems 3730/3730xl series data collection software | < 5.0.3 | 5.0.3 |
| applied biosystems seqstudio flex series instrument software | < 1.2.1 | 1.2.1 |
| applied biosystems seqstudio genetic analyzer data collection software | < 1.2.6 | 1.2.6 |
| genemapper id-x | < 1.7.4 | 1.7.4 |
The affected Thermo Fisher Applied Biosystems Genetic Analyzers are vulnerable because .fsa/.hid output files can be edited. An attacker could tamper with these files, altering DNA data and resulting in inaccurate DNA test outcomes.