| CVE-2023-29066 | BAIXO3.5 | The FACSChorus software does not properly assign data access privileges for operating system user accounts. A non-administrative OS account can modify information stored in the local application data folders. | 28/11/2023 |
| CVE-2023-29065 | MÉDIO4.3 | The FACSChorus software database can be accessed directly with the privileges of the currently logged-in user. A threat actor with physical access could potentially gain credentials, which could be used to alter or destroy data stored in the database. | 28/11/2023 |
| CVE-2023-29064 | MÉDIO4.3 | The FACSChorus software contains sensitive information stored in plaintext. A threat actor could gain hardcoded secrets used by the application, which include tokens and passwords for administrative accounts. | 28/11/2023 |
| CVE-2023-29063 | BAIXO2.4 | The FACSChorus workstation does not prevent physical access to its PCI express (PCIe) slots, which could allow a threat actor to insert a PCI card designed for memory capture. A threat actor can then isolate sensitive information such as a BitLocker encryption key from a dump of the workstation RAM during startup. | 28/11/2023 |
| CVE-2023-29062 | BAIXO3.8 | The Operating System hosting the FACSChorus application is configured to allow transmission of hashed user credentials upon user action without adequately validating the identity of the requested resource. This is possible through the use of LLMNR, MBT-NS, or MDNS and will result in NTLMv2 hashes being sent to a malici | 28/11/2023 |
| CVE-2023-29061 | MÉDIO5.2 | There is no BIOS password on the FACSChorus workstation. A threat actor with physical access to the workstation can potentially exploit this vulnerability to access the BIOS configuration and modify the drive boot order and BIOS pre-boot authentication. | 28/11/2023 |
| CVE-2023-29060 | MÉDIO5.7 | The FACSChorus workstation operating system does not restrict what devices can interact with its USB ports. If exploited, a threat actor with physical access to the workstation could gain access to system information and potentially exfiltrate data. | 28/11/2023 |
| CVE-2023-30565 | BAIXO3.5 | An insecure connection between Systems Manager and CQI Reporter application could expose infusion data to an attacker. | 13/07/2023 |
| CVE-2023-30564 | MÉDIO6.9 | Alaris Systems Manager does not perform input validation during the Device Import Function. | 13/07/2023 |
| CVE-2023-30563 | ALTO8.2 | A malicious file could be uploaded into a System Manager User Import Function resulting in a hijacked session. | 13/07/2023 |
| CVE-2023-30562 | MÉDIO6.7 | A GRE dataset file within Systems Manager can be tampered with and distributed to PCUs.
| 13/07/2023 |
| CVE-2022-47376 | ALTO7.3 | The Alaris Infusion Central software, versions 1.1 to 1.3.2, may contain a recoverable password after the installation. No patient health data is stored in the database, although some site installations may choose to store personal data. | 13/06/2023 |
| CVE-2022-30277 | MÉDIO5.7 | BD Synapsys™, versions 4.20, 4.20 SR1, and 4.30, contain an insufficient session expiration vulnerability. If exploited, threat actors may be able to access, modify or delete sensitive information, including electronic protected health information (ePHI), protected health information (PHI) and personally identifiable i | 02/06/2022 |
| CVE-2020-25165 | ALTO7.5 | BD Alaris PC Unit, Model 8015, Versions 9.33.1 and earlier and BD Alaris Systems Manager, Versions 4.33 and earlier The affected products are vulnerable to a network session authentication vulnerability within the authentication process between specified versions of the BD Alaris PC Unit and the BD Alaris Systems Manag | 13/11/2020 |
| CVE-2019-13517 | ALTO8.8 | In Pyxis ES Versions 1.3.4 through to 1.6.1 and Pyxis Enterprise Server, with Windows Server Versions 4.4 through 4.12, a vulnerability has been identified where existing access privileges are not restricted in coordination with the expiration of access based on active directory user account changes when the device is | 06/09/2019 |
| CVE-2018-10595 | MÉDIO6.3 | A vulnerability in ReadA version 1.1.0.2 and previous allows an authorized user with access to a privileged account on a BD Kiestra system (Kiestra TLA, Kiestra WCA, and InoqulA+ specimen processor) to issue SQL commands, which may result in loss or corruption of data. | 24/05/2018 |
| CVE-2018-10593 | MÉDIO5.6 | A vulnerability in DB Manager version 3.0.1.0 and previous and PerformA version 3.0.0.0 and previous allows an authorized user with access to a privileged account on a BD Kiestra system (Kiestra TLA, Kiestra WCA, and InoqulA+ specimen processor) to issue SQL commands, which may result in data corruption. | 24/05/2018 |
| CVE-2017-6022 | CRÍTICO9.8 | A hard-coded password issue was discovered in Becton, Dickinson and Company (BD) PerformA, Version 2.0.14.0 and prior versions, and KLA Journal Service, Version 1.0.51 and prior versions. They use hard-coded passwords to access the BD Kiestra Database, which could be leveraged to compromise the confidentiality of limit | 30/06/2017 |
| CVE-2016-9355 | MÉDIO5.3 | An issue was discovered in Becton, Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit, Version 9.5 and prior versions, and Version 9.7. An unauthorized user with physical access to an Alaris 8015 PC unit may be able to obtain unencrypted wireless network authentication credentials and other sensitive techni | 13/02/2017 |
| CVE-2016-8375 | MÉDIO4.9 | An issue was discovered in Becton, Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit, Version 9.5 and prior versions, and Version 9.7, and 8000 PC unit. An unauthorized user with physical access to an affected Alaris PC unit may be able to obtain unencrypted wireless network authentication credentials and | 13/02/2017 |